Feature | November 23, 2011 | Constantine A. Mantz, M.D.

One method for using SBRT for prostate cancer.

21st Century Oncology has treated more than 80 prostate cancer patients using a method of SBRT with a Varian linear accelerator and Calypso tumor localization system.


Stereotactic body radiation therapy (SBRT) is an investigational radiotherapy modality for the treatment of localized prostate cancer. Conventional radiotherapy options include prolonged schedules of daily treatment using relatively low radiation doses over a period of six to nine weeks. SBRT offers promise of an expedient, well-tolerated and effective schedule of only five treatments of larger doses for selected patients for whom current conventional treatment options would be either medically contraindicated or refused by the patient for reasons of inconvenience.

As an extremely hypofractionated therapy, SBRT furthermore provides a test of the radiobiologic principles that support fewer and larger fractions for prostate cancer irradiation. Accumulating radiobiological analyses of clinical outcomes suggest that the ?:? ratio of prostate cancer may be exceptionally low, with some studies reporting calculated ratios as low as 1.5. 1–8  If true, then prostate tumor effect may be highly sensitive to fraction size and perhaps even more so than the late effects on surrounding normal tissue. Therefore, the clinical implication of a hypofractionated treatment schedule for prostate cancer may be improved tumor control rates, while maintaining a similar biologically effective dose (BED) for normal tissue late effects, when compared to a conventionally fractionated regimen.

A fundamental technical demand of any prostate SBRT delivery system is its ability to respond to organ motion. Prostate movement is stochastic, meaning the gland may move in any direction at any time unpredictably. 9–10

In high dose rate (HDR) brachytherapy hypofractionation, organ motion is accounted for by placement of afterloading catheters into the prostate, and the catheters then move with the gland. For external modalities, implanted fiducial markers are commonly used to correct for interfractional organ motion prior to treatment delivery through the use of onboard imaging tools available on modern accelerators.  

However, this may be inadequate for hypofractionated therapy, as intrafractional prostate motion may also occur with displacements of up to 1 cm. 11–16  When a small number of large fractions are administered, there is an increased need to account for this source of localization error in order to maintain the minimal planning target volume (PTV) margins necessary for SBRT. Consequently, this reduces normal tissue complication probability and increases tumor control probability. 17–18  An optimal solution for external modalities is, therefore, a system that couples real-time tracking of the prostate to a linear accelerator and permits corrective action to changes in prostate position within a treatment fraction.

Real-Time Tumor Tracking
Varian Trilogy is a rapid output linear accelerator equipped with: (1) a multileaf collimator capable of producing complex and precise dose distributions; and (2) an integrated image guidance system. Briefly, the system provides image guidance by robotically extending an opposed kV X-ray source and amorphous silicon flat panel imager tandem over the target volume. The source-imager tandem may then acquire planar or volumetric imaging of implanted fiducial markers and/or the target to permit corrections of target setup error prior to treatment.  

During treatment – and particularly for larger fractions that require longer treatment times – imaging may need to be repeated to identify and correct intrafractional target motion. While technically possible, repeat imaging during a treatment fraction is time-consuming and is still unlikely to identify every instance of target excursion outside the treatment volume.  

The Calypso 4-D localization system is a real-time target tracking system that provides a more efficient solution to the problem of intrafractional target motion. This system consists of an electromagnetic array that can detect the positions of signal-emitting transponders implanted in the target and provides continuous information on target position with submillimeter accuracy.  

The system creates an electromagnetic field within the patient, which induces the implanted transponders to oscillate at a specific frequency. When the field is turned off, the transponders emit a signal that is received by the system to determine precisely their positions within the prostate. This function is performed 10 times per second to provide real-time target tracking. There is no use of X-rays; so no added radiation exposure for the patient.

Varian Trilogy, when coupled with the Calypso system, permits highly precise radiation dose shaping with real-time target tracking to account for intrafractional prostate motion during SBRT. As such, this combined system allows for the identification of potentially adverse movement of the prostate during treatment delivery and subsequent correction — either by stopping the treatment beam and waiting for the prostate to return to its normal treatment position or by repositioning the patient — in order to maximize the therapeutic effect of the treatment and mitigate its associated risks of rectal and bladder complications.  

Our Method
Our method of prostate SBRT with the Varian-Calypso system is as follows: The patient is set up on the treatment table with initial target localization accomplished through Calypso electromagnetic guidance. Varian’s onboard imager then generates a volumetric 3-D image of the prostate as verification of target geometry prior to delivery of a large fraction dose.

We emphasize this step as critical to the overall treatment process. We have observed among our study patients with some frequency a perfect match of fiducial markers to their reference coordinates, but deformation of the prostate on the same cone beam computed tomography (CT). When observed, this event usually occurs for a prostate that has a particularly concave interface with the rectum. With increased rectal filling, this concavity is exaggerated yet further, pushing the posterolateral prostate outside the field and bringing the midline anterior rectum further into the field. Consequently, the high-dose region misses a portion of the target and increasingly encroaches upon normal rectal tissue.  

In this circumstance, we have had the patient evacuate his rectum and then return to the treatment table for re-imaging. Once the set-up conditions are satisfied, the treatment can begin. The Calypso system is active at all times during beam delivery, and any identified transponder excursion beyond the 2 mm threshold we have assigned for corrective action stops the beam. As described earlier, either the beam is started again once the prostate is observed to have returned to its treatment position or the patient is repositioned and the entire set-up process is repeated. The treatment delivery then resumes and is completed.

To date, we have enrolled and treated more than 80 patients on study using this methodology since 2007. Our protocol calls for the delivery of 5 fractions of 8 Gy on an every-other-day schedule for a total dose of 40 Gy. Patients eligible for study are diagnosed with localized prostate cancer of low or intermediate grade and have a prostate gland measuring less than 60 cc on ultrasound volume assessment.  

We have not observed any serious toxicities requiring surgical correction or other procedures, and we have not observed any cancer recurrences. While highly encouraged by these favorable early results, we recognize that longer followup with a larger number of study patients is needed to provide further validation of these promising outcomes and support for prostate SBRT as an alternative to other radiation modalities in the management of low-risk prostate cancer.

Constantine Mantz, M.D., a board-certified radiation oncologist trained at University of Chicago, is chief medical officer at 21st Century Oncology. He has published scientific articles and presented original research at many international meetings on prostate cancer, breast cancer, head and neck cancer and lung cancer. He has participated in legislative and related efforts about payment methodologies and coverage decisions pertaining to radiation therapy services.

References:
1. Brenner DJ, Hall EJ. “Fractionation and protraction for radiotherapy of prostate carcinoma.” Int J Radiat Oncol Biol Phys 43:1095-1101; 1999.
2. Brenner DJ, Martinez AA, Edmundson GK, et al. “Direct evidence that prostate tumors show high sensitivity to fractionation (low ?/? ratio), similar to late-responding normal tissue.” Int J Radiat Oncol Biol Phys 52: 6-13; 2002.
3. Wang JZ, Guerrero M, Allen LX. “How low is the ?/? ratio for prostate cancer?” Int J Radiat Oncol Biol Phys 55: 194-203; 2003.
4. Fowler JF, Ritter MA, Chappell RJ, et al. “What hypofractionated protocols should be tested for prostate cancer?” Int J Radiat Oncol Biol Phys 56: 1093-1104; 2003.
5. King CR, Fowler JF. “A simple analytic derivation suggests that prostate alpha/beta ratio is low.” Int J Radiat Oncol Biol Phys 51: 213-214; 2001.
6. Fowler J, Chappell R, Ritter M. “Is ?/? for prostate tumors really low?” Int J Radiat Oncol Biol Phys 50: 1021-1031; 2001.
7. Brenner DJ. “Toward optimal external-beam fractionation for prostate cancer.” Int J Radiat Oncol Biol Phys 48: 315-316; 2000.
8. Duschesne GM, Peters LJ. “What is the ?/? ratio for prostate cancer? Rationale for hypofractionated high-dose-rate brachytherapy.” Int J Radiat Oncol Biol Phys 44: 747-748; 1999.
9. Byrne TE. “A review of prostate motion with considerations for the treatment of prostate cancer.” Med Dosim 30: 155-161; 2005.
10. Kitamura K, Shirato H, Steppenwoolde Y, et al. “Three-dimensional intrafractional movement of prostate measured during real-time tumor-tracking radiotherapy in supine and prone treatment positions.” Int J Rad Oncol Biol Phys 53: 1117-1123; 2002.
11. Britton KR, Takai Y, Mitsuya M, et al. “Evaluation of inter- and intra-fraction organ motion during intensity modulated radiation therapy (IMRT) for localized prostate cancer measured by a newly developed on-board image-guided system.” Radiat Med 23: 14-24; 2005.
12. Huang E, Dong L, Chandra A, et al. “Intrafraction prostate motion during IMRT for prostate cancer.” Int J Radiat Oncol Biol Phys 53: 261-268; 2002.
13. Nederveen AJ, van der Heide UA, Dehnad H, et al. “Measurements and clinical consequences of prostate motion during a radiotherapy fraction.” Int J Radiat Oncol Biol Phys 53: 206-214; 2002.
14. Hossain S, et al.  “Simulated real-time image guided intrafraction tracking-delivery for hypofractionated prostate IMRT.” Med Phys 35: 4041-4048; 2008.
15. Boike T, Anderson J, Novotny R, et al. “Intra-fraction and inter-fraction prostate motion associated with stereotactic body radiation therapy (SBRT) and image guided radiation therapy (IGRT).” Int J Radiat Oncol Biol Phys 69: S355; 2007.
16. Kupelian P, et al. “Multi-institutional clinical experience with the Calypso System in localization and continuous, real-time monitoring of the prostate gland during external radiotherapy.” Int J Radiat Oncol Biol Phys 67: 1088-1098; 2007.
17. Kitamura K, Shirato H, Shinohara K, et al. “Reduction in acute morbidity using hypofractionated intensity-modulated radiation therapy assisted with a fluoroscopic real-time tumor tracking system for prostate cancer: Preliminary results of a phase I/II study.” Cancer J  9: 268-276; 2003.
18. Cheung P, Sixel K, Morton G, et al. “Individualized planning target volumes for intrafraction motion during hypofractionated intensity-modulated radiotherapy boost for prostate cancer.” Int J Radiat Oncol Biol Phys 62: 418-425; 2005.


Related Content

News

Aug. 5, 2024 — Researchers from The University of Texas MD Anderson Cancer Center have demonstrated that adding ...

Time August 09, 2024
arrow
News | PET-CT

July 31, 2024 — In a head-to-head comparison with FDG PET/CT, FDG PET/MRI demonstrated comparable or superior diagnostic ...

Time July 31, 2024
arrow
News | Radiology Business

July 31, 2024 — The American Registry of Radiologic Technologists (ARRT) announced the three Registered Technologists (R ...

Time July 31, 2024
arrow
Feature | Radiation Oncology | By Christine Book

News emerging from several leading organizations and vendors in the radiation therapy arena came in at a fast pace in ...

Time July 30, 2024
arrow
News | Radiopharmaceuticals and Tracers

July 24, 2024 — Telix Pharmaceuticals Limited announced that the United States (U.S.) Food and Drug Administration (FDA) ...

Time July 24, 2024
arrow
News | Radiation Therapy

July 22, 2024 — RefleXion Medical, an external-beam theranostic oncology company, today announced that researchers from ...

Time July 22, 2024
arrow
News | ASTRO

July 18, 2024 — The members of the American Society for Radiation Oncology (ASTRO) recently elected five new officers to ...

Time July 18, 2024
arrow
News | PET-CT

July 16, 2024 — A new research paper was published in Oncotarget's Volume 15 on June 20, 2024, titled, “Comparison of ...

Time July 16, 2024
arrow
News | Digital Pathology

July 12, 2024 — AGFA HealthCare, a global leader in healthcare imaging management solutions, announced that Enterprise ...

Time July 12, 2024
arrow
News | Digital Pathology

July 12, 2024 — Diagnosing cancer and providing the personalized therapy it often requires, is a collaborative effort ...

Time July 12, 2024
arrow
Subscribe Now