July 16, 2008 - Earlier this week, the FDA released a notification that the X-rays used during a CT examination may cause some implanted and external electronic medical devices to malfunction.

The devices include cardiac pacemakers, defibrillators, neurostimulators, external drug infusion pumps, cochlear implants and retinal implants.

According to the FDA statement, this type of interference might be reported more frequently because of the increased utilization of CT, the higher dose-rate capability of newer CT machines, an increase in the number of patients with implanted and externally worn electronic medical devices and better reporting systems.

Among the adverse events reported include unintended “shocks” (i.e., stimuli) from neurostimulators, malfunctions of insulin infusion pumps and transient changes in pacemaker output pulse rate.

The FDA recommends that before a CT scan, the operator should use CT scout views to determine if implanted or externally worn electronic medical devices are present and if so, their location relative to the programmed scan range.

For CT procedures in which the medical device is in or immediately adjacent to the programmed scan range, the operator should:

-Determine the device type

-Try to move external devices out of the scan range, if practical

-Ask patients with neurostimulators to shut off the device temporarily while the scan is performed

-Minimize X-ray exposure to the implanted or externally worn electronic medical device by: using the lowest possible X-ray tube current consistent with obtaining the required image quality; and making sure that the X-ray beam does not dwell over the device for more than a few seconds.

The FDA requested clinicians submit reports of any adverse events related to the CT and implanted or external electronic devices via www.fda.gov/MedWatch/report.htm or by calling 1-800-FDA-1088.

For more information: www.fda.gov/cdrh/safety.html.


Related Content

News | Computed Tomography (CT)

At the annual AHRA (American Healthcare Radiology Administrators) conference in Orlando, Florida, Bayer announced an ...

Time August 09, 2024
arrow
Videos | Radiology Business

Find actionable insights to achieve sustainability and savings in radiology in this newest of ITN’s “One on One” video ...

Time July 30, 2024
arrow
News | Radiopharmaceuticals and Tracers

July 24, 2024 — Telix Pharmaceuticals Limited announced that the United States (U.S.) Food and Drug Administration (FDA) ...

Time July 24, 2024
arrow
News | RSNA

July 23, 2024 — Professional registration is open for RSNA 2024, the world’s largest radiology forum. This year’s theme ...

Time July 23, 2024
arrow
News | Artificial Intelligence

July 22, 2024 — Healthcare artificial intelligence (AI) systems provider, Qure.ai, has announced its receipt of a Class ...

Time July 22, 2024
arrow
News | PET-CT

July 16, 2024 — A new research paper was published in Oncotarget's Volume 15 on June 20, 2024, titled, “Comparison of ...

Time July 16, 2024
arrow
News | Prostate Cancer

July 11, 2024 — GE HealthCare’s MIM Software, a global provider of medical imaging analysis and artificial intelligence ...

Time July 11, 2024
arrow
News | Pediatric Imaging

June 25, 2024 — Rady Children’s Hospital-San Diego, one of the nation’s top pediatric health care systems, today ...

Time June 25, 2024
arrow
News | Artificial Intelligence

June 18, 2024 — The advancement of Artificial Intelligence (AI) in healthcare to support diagnostic decision making ...

Time June 18, 2024
arrow
News | Artificial Intelligence

June 5, 2024 — Nano-X Imaging, an innovative medical imaging technology company, today announced that its deep-learning ...

Time June 05, 2024
arrow
Subscribe Now